HME filter leakage pressure drop comprehensive tester is mainly used for anesthesia and breathing equipment humidification of human breathing gas heat and moisture exchanger air tightness test and pressure drop test, HME filter leakage pressure drop comprehensive tester equipment in strict accordance with ISO9360 standard manufacturing, the use of high-precision mass flow method to test the leakage of HME filter, and automatically complete the pressure drop measurement under 30L/min, 60L/min, 90L/min three kinds of flow. HEM filter leakage pressure drop comprehensive tester is made of materials without secondary pollution, ergonomic design, simple operation, accurate testing and basic intelligent judgment function, suitable for HME filter manual production line and automated production line.
Main features
Test mode: automatic measurement and manual measurement can be switched
Test items: air leakage test + resistance test
CP mode: HeM filter leakage is tested in constant pressure mode
CF mode: Pressure drop measurement in constant flow mode (simulated air resistance test)
Intelligent judgment: leakage test pass pass pass alarm, pressure drop test pass pass alarm
Product material: food grade stainless steel manufacturing, high-precision filter to ensure that the product is pollution-free
Complies with the standard: HME filter leakage pressure drop comprehensive tester
ISO5356-1, YY/T 1040.1 Cone fittings for anesthesia and respiratory devices
ISO9360.1, YY/T0735.1 Anesthesia and respiratory equipment Heat and moisture exchangers (HME) that humidify human breathing gases
ISO9360.2, YY/T0735.2 Anesthesia and respiratory equipment Heat and moisture exchangers (HMEs) for humidifying human breathing gases
Technical parameters of HME filter leakage pressure drop comprehensive tester
Two-in-one measurement items: |
High-precision flow measurement method, flow resistance test HME filter leakage pressure drop comprehensive tester |
|
Automatic measuring circuit: |
The instrument automatically completes the leak test + pressure drop test (resistance test) |
|
Intelligent judgment function: |
Leak test qualified and unqualified alarm, pressure drop test qualified and unqualified alarm |
|
CP mode |
Test pressure range |
Inflation pressure: 1-999.9cmH2O & 1-98.9kPa adjustable |
Leak Test Judgment |
Leakage measurement range: 0.5-200ml/min |
|
CF mode |
Flow setting range |
Flow selection value: 30L/min, 60Lmin, 90L/min or 100Lmin |
Resistance test range |
Resistance test range: 0-100 mmH2O/min |
|
Resistance determination range |
When constant current is 30L/min: resistance value ≤35Pa (can be set) |
|
Time setting range: |
Delay time: 0.1-999.9sec |
|
safety protection |
The device man-machine interface sets a password to prevent non-professionals from misoperation. |
|
test interface |
YY1040 inner cone interface |
|
equipment noise |
Distance 1m≤75dB |
|
Use environment |
0~40°C/20~85%RH (non-condensing) |
|
Electric Source |
AC 90~240V 50/60Hz |
|
Gas Source |
>0.4MPa stable and clean compressed air |
|
Air supply interface |
Φ8mm trachea |
|
Performance Rate |
about 50W |