Product performance
Inflation method: large-diameter rapid inflation to achieve rapid testing of long breathing tubes
Parameter setting: inflation time, pressure holding time, test time and judgment standard can be set
Judgment function: intelligent and rapid judgment of qualified and unqualified and sound and light alarm prompt
Real-time display: real-time display of inflation pressure, test results, test curves and judgment results
NG product self-locking: unqualified products are automatically locked and unlocked by engineers to avoid error outflow
Data storage: data statistics are stored in the U disk, which is convenient for data export to EXCEL for analysis
Automatic start: realize buttonless start function, the test can be automatically started when the product is loaded
Standards Compliant
ISO5356-1, YY/T 1040.1 Conical joints for anesthesia and respiratory equipment
ISO5367, YY/T0461 breathing circuit for anesthesia machines and ventilators
ISO9360.2, YY/T0735.2 Anesthesia and respiratory equipment heat and moisture exchanger for humidifying human breathing gas
Model Specifications
|
ATC315S-2
|
measurement method
|
High precision mass flow method
|
Measurement channel
|
Single channel independent measurement
|
Voltage regulation range
|
Inflation pressure: 1-50kPa adjustable
Resolution: 0.01Kpa
Display accuracy: 0.25% F.S
|
Test unit
|
ml/min, kPa, cmH2O
|
Leakage measurement range
|
Leak test range: 0-500ml/min
Leakage resolution: 0.01ml/min
Test accuracy: 0.5%F.S
|
Time setting range
|
Delay time: 0.1-999.9sec
Inflation time: 0.1-999.9sec
Equilibration time: 0.1-999.9sec
Detection time: 0.1-999.9sec
Exhaust time: 0.1-999.9sec
|
Pre-inflation function
|
The pre-inflated function of the large aperture makes the measurement faster
|
Auto start
|
Automatic starting pressure can be set according to product volume
|
Judgment function
|
Qualified and unqualified automatic judgment
|
Lock function
|
Unqualified products are automatically locked
|
External control input
|
1 start, 1 stop, 1 hold
|
External control output
|
1 channel qualified alarm, 1 channel unqualified alarm, 1 group of cylinder control
|
Accessibility
|
bubble function, hold function
|
Data save
|
Automatically record test data USB dump read analysis
|
safety protection
|
Set password protection to prevent misuse by non-professionals
|
Equipment noise
|
Distance 1 meter≤75dB
|
Use environment
|
0~40°C/20~85%RH (non-condensing)
|
Electric Source
|
AC 90~240V 50/60Hz
|
Gas Source
|
>0.4MPa stable and clean compressed air
|
Air supply interface
|
Φ8mm trachea quick interface
|
Performance Rate
|
about 50W
|
The whole machine meets the production process requirements of the production line and the requirements of the medical production environment, with simple operation and convenient maintenance.
When the product volume and test conditions are different, please choose a different test range or call Dongguan Otto Company at 4006016628. Our company will recommend a reasonable test plan to meet your HMEF breathing circuit leak test, HMEs breathing filter leak test and other related Leakage, flow and resistance testing of medical products.